Injection: MicroSet. Watch video.
Incision size: greater or equal to 1.8mm
Filtration: UV and blue light blocker
Optic: biconvexe aspheric aberration-correcting
Material: 25% hydrophilic acrylic
Optic body diameter: 6.15mm (6.3mm from -10D to -1D)
Overall diameter: 10.75mm (11mm from -10D to -1D)
Angulation:
Power: from -10D to +10D (1D increments), from +10D to +30D (0.5D increments), from +30D to +35D (1D increments)
Suggested constants for IOL power calculation:

Haigis formula: a0=1.36 (IOL Master) 1.04 (US), a1=0.4, a2=0.1
Hoffer Q: pACD=5.52 (IOL Master) 5.26 (US)
Holladay 1: Sf=1.74 (IOL Master) 1.48 (US)
SRK II: A =119.3 (IOL Master) 118.9 (US)
SRK/T: A =118.9 (IOL Master) 118.6 (US)

 

 

 

 

 

Improving quality of vision

The Micro AY aspheric optic  partly compensates the corneal spherical aberration for improving contrast sensitivity while preserving depth of field.
Learn more

 

Ensuring retinal protection

The Micro AY Blue Tech blocker optimises UV light filtration and attenuates potentially phototoxic blue light.
Learn more

 

Lowering Yag rates

The clinical efficiency of the 2-Step Technology 360° square edge was demonstrated 2 years after implantation. Minimal Yag rates were found in various studies.